SIRA – UV-Analyzer
Description
The SIRA process gas analyzer is the standard device for the prescribed determination of the odorant concentration in accordance with DVGW worksheet G 280 and G24. The analyzer is also able to determine the concentration of aromatic hydrocarbons.
The measurement is carried out using UV spectrometry. Adapted analysis systems including internal zero gas generation are available for the respective odorant. The analysis system can be adjusted at any time when the odorant is changed.
The analyzer is suitable for operation in natural gas and hydrogen. The measurement of aromatic hydrocarbons in nitrogen is also possible.
Measurements with different gases can be carried out time-controlled through optimal switching of the measuring points.
Application
- Biogas
- Sewage gas plants
- Landfill sites
- Oil and Gas
- Chemistry
Measured quantities
Substance | measuring range | error limits |
---|---|---|
Tetrahydrothiophen (THT) | 0 – 50 mg/m³ | 0,75 mg/m³ or 10 % of the measured value (the greater value applies) |
Gasodor® S-Free | 0 – 50 mg/m³ | 0,75 mg/m³ or 10 % of the measured value (the greater value applies) |
Benzene | 0 – 50 mg/m³ | 0,75 mg/m³ or 10 % of the measured value (the greater value applies) |
Toluene | 0 – 50 mg/m³ | 0,75 mg/m³ or 10 % of the measured value (the greater value applies) |
Ethylbenzene | 0 – 50 mg/m³ | 0,75 mg/m³ or 10 % of the measured value (the greater value applies) |
Xylene | 0 – 50 mg/m³ | 0,75 mg/m³ or 10 % of the measured value (the greater value applies) |